1、Agilent 6400 SeriesTriple Quad LC/MSSystemInstallation GuideNotices© Agilent Technologies, Inc. 2006-2010No part of this manual may be reproduced inany form or by any means (including elec-tronic storage and retrieval or translationinto a foreign language without prior agree-ment and written co
2、nsent from AgilentTechnologies, Inc. as governed by UnitedStates and international copyright laws.Manual Part NumberG3335-90081EditionTenth Edition, May 2010Printed in USAAgilent Technologies, Inc.5301 Stevens Creek Blvd.Santa Clara, CA 95051Warranty The material contained in this docu-ment is provi
3、ded “as is,” and is sub-ject to being changed, without notice, in future editions. Further, to the max-imum extent permitted by applicable law, Agilent disclaims all warranties, either express or implied, with regard to this manual and any information contained herein, including but not limited to t
4、he implied warranties of merchantability and fitness for a par-ticular purpose. Agilent shall not be liable for errors or for incidental or consequential damages in connec-tion with the furnishing, use, or per-formance of this document or of any information contained herein. Should Agilent and the u
5、ser have a separate written agreement with warranty terms covering the material in this document that conflict with these terms, the warranty terms in the sep-arate agreement shall control.Technology Licenses The hardware and/or software described in this document are furnished under a license and m
6、ay be used or copied only in accor-dance with the terms of such license.Restricted Rights LegendU.S. Government Restricted Rights. Soft-ware and technical data rights granted tothe federal government include only thoserights customarily provided to end user cus-tomers. Agilent provides this customar
7、ycommercial license in Software and techni-cal data pursuant to FAR 12.211 (TechnicalData and 12.212 (Computer Software and,for the Department of Defense, DFARS252.227-7015 (Technical Data - CommercialItems and DFARS 227.7202-3 (Rights inCommercial Computer Software or Com-puter Software Documentati
8、on.Safety NoticesCAUTION A CAUTION notice denotes a haz-ard. It calls attention to an operat-ing procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicate
9、d conditions are fully understood and met. WARNING A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly per-formed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indica
10、ted condi-tions are fully understood and met.Windows ® and Microsoft ® is a U.S. regis-tered trademark of Microsoft Corporation.Windows NT ® is a U.S registered trade-mark of Microsoft Corporation.Software RevisionThis guide is valid for the B.03.01 revision orhigher of the Agilent Ma
11、ssHunter Worksta-tion Software - Data Acquisition programand compatible Qualitative Analysis andQuantitative Analysis programs, until super-seded.In This GuideThis guide contains information to install the Agilent 6400 Series Triple Quad LC/MS System.1Instrument InstallationThis chapter describes th
12、e steps that are needed to install the Triple Quad instrument.2Software Installation and ConfigurationThis chapter describes the steps that are needed to install the MassHunter software.3Installation Completion and System VerificationI n this chapter, you finish the hardware setup.4Installation and
13、Verification of Other SourcesThis chapter contains the steps needed to install and verify the operation of the APCI and APPI sources.5Other Setup TasksThis chapter contains tasks that are related to setting up your Triple Quad instrument.6ReferenceThis chapter contains parameters for verifying perfo
14、rmance.Content1Instrument Installation7Step 1. Unpack and set up the Triple Quad instrument8Step 2. Prepare to install the system9Step 3. Set up the rough pump13Step 4. Connect the Triple Quad instrument16Step 5. Install the source23Step 6. Install the Agilent 1100 or 1200 series LC system292Softwar
15、e Installation and Configuration31Step 1. Set up the computer system32Step 2. Check PC network card configuration33Step 3. Prepare for installation35Step 4. Install the Data Acquisition program36Step 5. Install the Qualitative Analysis program41Step 6. Install the Quantitative Analysis program43Step
16、 7. Install and configure Microsoft Excel 200747Step 8. Install Service Packs for Data Acquisition51Step 9. Install Quantitative Analysis Reporting52Step 10. Check Excel Add-Ins for MassHunter54Step 11. Configure the instrument55Step 12. Copy the support folder to the computer56Step 13. Confirm the
17、LC firmware revision58Step 14. Finish the installation593Installation Completion and System Verification61Step 1. Condition the 1100 / 1200 LC62Step 2. Prepare the performance evaluation samples64Step 3. Review methods and run worklist66Step 4. Verify the Triple Quad sensitivity73To reset the Collis
18、ion Cell gas flow rate776Agilent 6400 Series Triple Quad LC/MS System Installation GuideContents4Installation and Verification of Other Sources81G1947B APCI Source82To install the G1947B APCI source82To verify the operation of the G1947B APCI source83G1978B Multimode Source84To install the G1978B mu
19、ltimode source84To verify the operation of the G1978B multimode source85G1971B APPI Source86To install the G1971B APPI source86To verify the operation of the G1971B APPI source87HPLC-Chip/MS Cube Interface88Step 1. Prepare performance verification sample88Step 2. Prepare the HPLC-Chip/MS90Step 3. An
20、alyze the results935Other Setup Tasks97To start the Data Acquisition program98To stop MassHunter Engines98To stop MassHunter processes in case of unrecoverable errors98To view the log book98To configure the network interface card99To confirm configuration of NICs100To remove the MassHunter Workstati
21、on software102To reconfigure the instrument1036Reference 1056460/6430 Reserpine Checkout Method Acquisition Parameters1067Agilent 6400 Series Triple Quad LC/MS SystemInstallation Guide Agilent Technologies1Instrument InstallationStep 1. Unpack and set up the Triple Quad instrument8Step 2. Prepare to
22、 install the system9Step 3. Set up the rough pump13Step 4. Connect the Triple Quad instrument16Step 5. Install the source23Step 6. Install the Agilent 1100 or 1200 series LC system29This chapter describes the steps that are needed to install the Triple Quad instrument. Figure 1 6460 Triple Quad LC/M
23、S8Agilent 6400 Series Triple Quad LC/MS System Installation Guide1Instrument InstallationStep 1. Unpack and set up the Triple Quad instrumentStep 1. Unpack and set up the Triple Quad instrument1Carefully examine all containers for external signs of damage. If damage is discovered, immediately contac
24、t the carrier and the Agilent Service Management.Retain shipping containers and material until contents are checked for completeness and instrument performance is verified.2Check off each item on the packing list, and verify the serial numbers.3Record the serial numbers in the installation documenta
25、tion. 4Report any discrepancies to Agilent Technologies.5Check that site preparation is complete, which includes power and gas supplies and chemical supplies needed for installation and performance verification.6Check the line voltage.Verify the proper line voltage and outlet ratings according to th
26、e instructions given in the Site Preparation Guide.NOTE Do not open the shipping containers until an Agilent representative is present to verify thecontents of each carton. Warranty claims for missing items are honored only if an Agilent representative is on site to verify the contents of each shipp
27、ing container as it is unpacked.Checkout and tuning samples are shipped separately to comply with safety regulations and to ensure prompt delivery. Make sure that these samples have been stored correctly, as indicated in the instructions in the box.Agilent 6400 Series Triple Quad LC/MS System Instal
28、lation Guide9Instrument Installation1Step 2. Prepare to install the systemStep 2. Prepare to install the systemBefore you begin, check that you have these parts:Utility knife and needle nose pliers1Unpack and set up the Triple Quad instrument.a Cut the tie wraps around the package. See Figure 2. Fig
29、ure 2Triple Quad instrument shipping containerb Lift the boxes off the top of the instrument. See Figure 3.10Agilent 6400 Series Triple Quad LC/MS System Installation Guide1 Instrument InstallationStep 2. Prepare to install the systemFigure 3Instrument unboxed on shipping pallet (6460 picturedc Remo
30、ve the front, top, right, back-left, and lower front-left cosmetic covers so that you can remove the shipping brackets. You do not need to remove the upper front-left cover. To remove the top cover, remove the two screws at the back side of the top cover that secure the back of the top cover to the
31、instrument chassis (see Figure 4. These screws are only needed when shipping the instrument. You do not need to reinstall the screws when you reinstall the top cover.Instrument Installation1Step 2. Prepare to install the system Figure 4Remove shipping screws from back of top panel. The screws to rem
32、ove are circled in this picture.d Remove the shipping brackets (a total of 7. You will need to use a 13-mm wrench. Figure 5Triple Quad on shipping pallet with covers removed (6460 pictured1Instrument InstallationStep 2. Prepare to install the systeme Take an inventory of the contents of the packages
33、.f Record serial numbers. The serial number label for the Triple Quad instrument is located on the lower left corner to the left of the instrument power push button.2Use four people, one at each corner of the instrument, to lift the Triple Quad instrument onto a stable bench or table.NOTE Do NOT lif
34、t from under the power switch bracket on the front left corner of the instrument.The bracket will bend. See Figure 6. Figure 6Do not lift from under power switch bracketWARNING Do not lift the Triple Quad without assistance. The instrument weighs up to 115 kg(255 lbs. Lift with your knees, not with
35、your back. Keep your back straight while you bend your knees.Instrument Installation1Step 3. Set up the rough pumpStep 3. Set up the rough pump1Unpack the mist filter, exhaust hose, and pump oil drip pan from the shipping kit.2Remove the rough pump from its shipping container:a Cut the straps that s
36、ecure the cardboard box to the wooden pallet.b Lift off the cardboard box.c Set aside the two bottles of pump oil and funnel.d Set aside the oil return kit. The felt plugs and screen inside the packaging are not needed and may be discarded.e Use a pair of needle nose pliers to remove the metal clips
37、 that secure the pump to the wooden pallet.3Set the pump into the pump oil drip pan.The pan is used to catch any small amounts of oil that may seep out of the pump. The pan can also contain all of the oil in the pump in case of a catastrophic seal failure.4Fill the rough pump with Inland 45 fluid.Th
38、e rough pump is drained of vacuum pump oil prior to shipment. Remove the oil fill cap and use the funnel to add Inland 45 pump fluid until the level in the site glass is approximately 2/3 full.5I nstall the oil mist filter.Remove the plugged outlet connector and install the new KF-25 fitting that co
39、mes with the pump. Do not use the outlet nozzle as described in the Edwards manual and on the warning tag. Install the mist filter on the outlet port KF-25 fitting. See figure on the next page.Make sure the mist filter drain plug is facing the rough pump ballast valve (toward the motor end of the pu
40、mp.6I nstall the oil return line:a Remove the drain plug from the mist filter and install the drain adapter.b I nstall the brass restrictor into the black oil-return tubing.c Connect the oil-return tubing between the rough pump ballast port fitting and the oil mist filter drain adapter. Make sure th
41、e end of the tubing with the brass restrictor is installed at the ballast port fitting.1Instrument InstallationStep 3. Set up the rough pumpd Use the plastic hose clips and a pair of needle nose pliers to secure the oil-return tubing. 7Check that the ballast valve is closed.Closing the ballast valve
42、 prevents additional air from being introduced to the pump. This allows the rough pump to reach a lower operating pressure.Figure 7Example rough pump. Note that your pump may look slightly different.Oil mist filterExhaust hoseRough pump hoseOil return lineBallast valveOil drain plugOil level sight g
43、lassOil drip panCAUTION The ballast valve must be opened (fully counterclockwise periodically to allow oiltrapped in the mist filter to return to the pump. If the oil is not returned to the pump, the pump will eventually run dry and be damaged.Instrument Installation1Step 3. Set up the rough pump8Fi
44、nd the Low Profile Exhaust Adaptor (p/n G1960-20003 and the O-ring (p/n 0905-1592. Put the O-ring on the exhaust adaptor.9Locate one KF-25 swing clamp (p/n 0100-1398, open the swing clamp, and use it to clamp the exhaust adaptor onto the oil mist filter.Connect the 1/2-inch I.D. Tygon exhaust tubing
45、 to the exhaust adaptor.16Agilent 6400 Series Triple Quad LC/MS System Installation Guide1Instrument InstallationStep 4. Connect the Triple Quad instrumentStep 4. Connect the Triple Quad instrument1Connect the rough hose (p/n G1969-20123 that is supplied in the ship kit to the KF-25 rough pump hose
46、connection on the back left of the instrument.2Connect the rough pump hose to the KF-25 inlet adapter on the rough pump.3Use the supplied hook/loop fastener tape to secure the drain bottle into a secondary containment tub.4Connect the spray chamber drain tube to the 1-inch fitting on the drain bottl
47、e. The drain tube can also be routed out the front of the Triple Quad if needed.5Attach the calibrant delivery system (CDS/inlet module waste tubing from the solvent selection valve on the Triple Quad to the ¼-inch fitting on the drain bottle.6Connect a length of the ½-inch Tygon tubing to
48、 the ½-inch fitting on the drain bottle, then connect the other end of the tubing to a vent connection that is separate from the vent used for the rough pump.CAUTION The drain bottle vent must be located away from the rough pump vent to prevent the rough pump exhaust from contaminating the Trip
49、le Quad spray chamber.Agilent 6400 Series Triple Quad LC/MS System Installation Guide17Instrument Installation1Step 4. Connect the Triple Quad instrumentFigure 8 Drain bottleSpray chamber drain tubeDrain bottle ventCDS inlet and LC waste fittingWaste drain bottleSecondary containment(customer suppli
50、ed18Agilent 6400 Series Triple Quad LC/MS System Installation Guide1Instrument InstallationStep 4. Connect the Triple Quad instrument7I nstall the gas filter:For 6410/6430:a Connect an appropriate length of the 1/4-inch teflon gas tubing (G1946-80078 to nitrogen gas source. Use the supplied nut and
51、ferrule kit (p/n 5183-0392. Keep some tubing for step c and step d . b Attach the other end of the of the tubing (from the nitrogen gas source to the BHT-4 gas filter. Use the nuts and ferrules supplied with the BHT-4 gas filter.c Use a piece of 1/4-inch teflon gas tubing to connect the BMT-4 gas fi
52、lter to the BHT-4 gas filter. Use the nuts and ferrules supplied with the BMT-4 and BHT-4 gas filters.d Connect a new length of 1/4-inch teflon gas tubing to the other end of the BHT-4 gas filter. Use the other nut and ferrule kit supplied with the BHT-4 gas filter.e Connect the gas tubing from the
53、BHT-4 gas filter (from step d to the nitrogen gas fitting on the back of the LC/MS system. Use the supplied nut and ferrule kit (p/n 5183-0392. f Secure the gas filter in a vertical position with tie wraps.This configures the filters in series with the supply tubing. See Figure 9 as an example for a
54、ssembling the gas trap in the correct configuration. Outlet to LC/MSBHT-4BMT-4Figure 9Gas filter con-nected in seriesInlet from nitrogen gas sourceAgilent 6400 Series Triple Quad LC/MS System Installation Guide19Instrument Installation1Step 4. Connect the Triple Quad instrumentFor 6460:a Locate the
55、2 1/4-inch brass T-unions(0100-0088 and loosen the brass nuts onthe T-unions.b Use an appropriate length of the 1/4-inchteflon gas tubing (G1946-80078 toconnect the nitrogen gas source to thecenter connector of one T-union.c Connect two 10-cm lengths of 1/4-inchteflon gas tubing to each side of theT
56、-union. Use the nuts and ferrulessupplied with the T-union.d Attach the other end of the 10-cm lengthsof 1/4-inch teflon gas tubings (fromstep c to the two BMT-4 gas filters. Usethe nuts and ferrules supplied with theBMT-4 gas filters.e Use a small length of gas tubings toconnect each BMT-4 gas filt
57、er to a BHT-4gas filter. Use the nuts and ferrules fromthe BMT-4 and BHT-4 gas filters.f Attach the other end of each BHT-4 gasfilter to a 10-cm length of 1/4-inc teflongas tubing. Use the nuts and ferrulessupplied with the BHT-4 gas filters.g Connect the other end of each 10-cm length of gas tubing
58、 (from step f to the ends of the remaining T-union. Use the nuts and ferrules supplied with the T-union.h Connect the gas tubing from the BHT-4 gas filter (from step g to the nitrogen gas fitting on the back of the LC/MS system. Use the supplied nut and ferrule kit (p/n 5183-0393.i Secure the gas filter in a vertical position with tie wraps.Outlet to LC/MSBHT-4BMT-4Figure 10Gas filter, con-nected first in series, then in parallel.Inlet from nitrogen gas so
溫馨提示
- 1. 本站所有資源如無(wú)特殊說(shuō)明,都需要本地電腦安裝OFFICE2007和PDF閱讀器。圖紙軟件為CAD,CAXA,PROE,UG,SolidWorks等.壓縮文件請(qǐng)下載最新的WinRAR軟件解壓。
- 2. 本站的文檔不包含任何第三方提供的附件圖紙等,如果需要附件,請(qǐng)聯(lián)系上傳者。文件的所有權(quán)益歸上傳用戶所有。
- 3. 本站RAR壓縮包中若帶圖紙,網(wǎng)頁(yè)內(nèi)容里面會(huì)有圖紙預(yù)覽,若沒(méi)有圖紙預(yù)覽就沒(méi)有圖紙。
- 4. 未經(jīng)權(quán)益所有人同意不得將文件中的內(nèi)容挪作商業(yè)或盈利用途。
- 5. 人人文庫(kù)網(wǎng)僅提供信息存儲(chǔ)空間,僅對(duì)用戶上傳內(nèi)容的表現(xiàn)方式做保護(hù)處理,對(duì)用戶上傳分享的文檔內(nèi)容本身不做任何修改或編輯,并不能對(duì)任何下載內(nèi)容負(fù)責(zé)。
- 6. 下載文件中如有侵權(quán)或不適當(dāng)內(nèi)容,請(qǐng)與我們聯(lián)系,我們立即糾正。
- 7. 本站不保證下載資源的準(zhǔn)確性、安全性和完整性, 同時(shí)也不承擔(dān)用戶因使用這些下載資源對(duì)自己和他人造成任何形式的傷害或損失。
最新文檔
- 房地產(chǎn)守價(jià)議價(jià)及SP配合培訓(xùn)
- 風(fēng)電技能培訓(xùn)課件圖片素材
- 各種護(hù)理導(dǎo)管防滑脫措施
- 小學(xué)教導(dǎo)處常規(guī)管理匯報(bào)
- 肺炎的公休座談會(huì)
- 頸椎病健康教育課件
- 領(lǐng)航職業(yè)英語(yǔ)課件下載
- 預(yù)防踩踏班會(huì)課件
- 崗前培訓(xùn)結(jié)業(yè)匯報(bào)
- 預(yù)防小學(xué)生溺水課件
- 國(guó)家開(kāi)放大學(xué)電大??啤堵眯猩缃?jīng)營(yíng)管理》2025期末試題及答案(試卷號(hào)2312)
- 《義務(wù)教育數(shù)學(xué)課程標(biāo)準(zhǔn)(2022年版)》解讀
- LY/T 2773-2016綠地月季栽培養(yǎng)護(hù)技術(shù)規(guī)程
- GB/T 7092-1993半導(dǎo)體集成電路外形尺寸
- GB/T 15909-2017電子工業(yè)用氣體硅烷
- 第一章有理數(shù)單元測(cè)試 人教版七年級(jí)數(shù)學(xué)上冊(cè)
- 門診病歷書寫基本規(guī)范-課件
- 做好新形勢(shì)下群眾工作培訓(xùn)課件
- FMEA(失效模式分析)教程課件
- PEP小學(xué)英語(yǔ)單詞分類匯總(含音標(biāo))
- NB∕T 10731-2021 煤礦井下防水密閉墻設(shè)計(jì)施工及驗(yàn)收規(guī)范
評(píng)論
0/150
提交評(píng)論